| Model Number |
4965-35 |
| Medical Device Problem Codes |
Failure to Capture (1081); Over-Sensing (1438); Pacing Problem (1439)
|
| Health Effect - Clinical Code |
Chest Pain (1776)
|
| Date of Event |
03/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient experienced chest pain.It was reported that the right atrial (ra) lead exhibited possible oversensing and loss of capture along with far field r-wave (ffrw) oversensing on stored electrograms (egm) resulting in false termination of atrial tachycardia/atrial fibrillation (at/af) episodes and inappropriate pacing.The lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 4965-35 lead was implanted on (b)(6) 2017.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported that the chest pain was not related to the ra lead.
|
| |
|