| Model Number |
XTACK-235-H |
| Medical Device Problem Codes |
Break (1069); Difficult to Remove (1528); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block h6: impact code f1001 is being used to capture the reportable event of aborted/cancelled procedure with a patient with unknown sedation status.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an x-tack endoscope helix tacking system (235cm) was used during a colon hemostasis procedure performed on (b)(6) 2024.During the procedure, the x-tack "presented a defect" when removing the catheter from the working channel, pulling the wire.The procedure was not completed.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an x-tack endoscope helix tacking system (235cm) was used during a colon hemostasis procedure performed on (b)(6) 2024.During the procedure, the x-tack "presented a defect" when removing the catheter from the working channel, pulling the wire.The procedure was not completed.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: an x tact was returned to boston scientific corporation where it was observed that the tack was stretched.It was not possible to replicate the event as it occurred during the procedure.The event of suture break and difficult to remove could not be confirmed.This investigation is assigned a most probable conclusion code of adverse event related to procedure.This conclusion was selected because during the product analysis the device returned with the first helix tack stretched.Most likely procedural factors such as lesion characteristics handling of the device the technique used by the physician (force applied) could have resulted in the damages encountered in the device.Block h6: impact code f1001 is being used to capture the reportable event of aborted cancelled procedure with a patient with unknown sedation status.
|
| |
|