| Model Number |
M00542430 |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Note: this report pertains to one of two devices used during the same procedure.Refer to manufacturer report number 3005099803-2024-01035 for the exalt model d scope.It was reported that and exalt model d controller was used for an endoscopic retrograde cholangiopancreatography (ercp) procedure on (b)(6) 2024.During procedure, the image was lost and was unable to be recovered after multiple attempts of powering on.The controller would not power on.The account switched to a reusable scope to complete the procedure.There were no patient complications related to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: device code a06 captures the reportable event of loss of visualization inside the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: device code a06 captures the reportable event of loss of visualization inside the patient.Block h10: with all the available information, boston scientific concludes that the reported allegation of loss of visualization was unable to be confirmed.Upon receipt of this device, the controller was thoroughly analyzed.Visual analysis identified finish damage to the top cover, front panel, and bottom of housing.During functional testing, the device performed as intended.The reported observation was not able to be duplicated or confirmed.Based on all gathered information, the probable cause selected for loss of visualization is no problem detected, which indicates that the device problem was not confirmed.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
|
| |
|
Event or Problem Description
|
|
Note: this report pertains to one of two devices used during the same procedure.Refer to manufacturer report number 3005099803-2024-01035 for the exalt model d scope.It was reported that and exalt model d controller was used for an endoscopic retrograde cholangiopancreatography (ercp) procedure on (b)(6) 2024.During procedure, the image was lost and was unable to be recovered after multiple attempts of powering on.The controller would not power on.The account switched to a reusable scope to complete the procedure.There were no patient complications related to this event.
|
| |
|