ETHICON ENDO-SURGERY, LLC. LIGACLIP* APPLIER W/MECH STOP; APPLIER, SURGICAL, CLIP Back to Search Results
Catalog Number LX107
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure, the clips are not capturing properly on the tissue.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 3/20/2024 d4: batch # unk additional information was requested and the following was obtained: "is the current patient status known? ¿ patient is safe as informed by or incharge.-was there any patient consequence or change in the post-operative care of the patient as a result of the event? (extended hospital stay, readmission, re-operation, etc.)- no" attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4), date sent: 5/23/2024.D4: batch # unk.Investigation summary the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the lx107 device was returned nonfunctional as the jaws were misaligned; therefore, the clips could not be loaded into the jaws.The event reported was confirmed and it is related to improper use of the device.Please note that as stated in the ifu: "all surgical instruments are subject to a degree of wear and tear because of normal use.A regular and precise visual check of the instrument should be made before each use.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key18941540
Report Number3005075853-2024-02278
Device Sequence Number8232097
Product Code GDO
UDI-Device Identifier10705036012856
UDI-Public10705036012856
Combination Product (Y/N)N
Initial Reporter CountryMU
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberLX107
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/06/2024
Supplement Date Received by Manufacturer05/23/2024
Initial Report FDA Received Date03/20/2024
Supplement Report FDA Received Date05/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-