| Model Number |
G18629 |
| Medical Device Problem Codes |
Break (1069); Device Damaged Prior to Use (2284)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown or unavailable.E1: phone - (b)(6).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
As reported, before an unspecified procedure, the tip of the roadrunner nimble hydrophilic wire guide separated.The device did not make patient contact.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention and did not experience any adverse effects due to this occurrence.Additional information regarding event details, patient anatomy and outcome has been requested but is not available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d9, h3: device evaluated by mfg = other (81) - device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Additional information was received 27mar2024.The wire guide holder was not flushed prior to attempted removal of the wire guide.The procedure was completed with an unspecified wire guide.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation ¿ evaluation: as reported, before an unspecified procedure, the tip of the roadrunner nimble hydrophilic wire guide separated.The device did not make patient contact.The wire guide holder was not flushed prior to attempted removal of the wire guide.The procedure was completed with an unspecified wire guide.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention and did not experience any adverse effects due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), quality control, and instructions for use (ifu), as well as a visual inspection, of the returned device, were conducted during the investigation.One roadrunner nimble hydrophilic wire guide was returned to cook for evaluation in an open package with label.A visual examination noted the tip of the wire guide was broken and the coil of the wire guide was exposed.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot revealed no recorded non-conformances relevant to the failure mode.A database search did not identify any other events associated with the reported device lot.The evidence from the complaint file, device history record, complaint history and quality control documents indicate that the complaint device was manufactured to specification as well as other items in the lot or similar devices in the field or in house.Cook also reviewed product labeling.The ifu supplied with the device (tocwg_rev1) was reviewed and includes the following instructions for activating hydrophilic coating: the hydrophilic coating on the wire guide is activated by immersion in sterile water or sterile saline solution.1.Prior to using the wire guide, fill a syringe with sterile water or sterile saline solution and attach it to the flushing port on the wire guide.2.Inject enough solution to wet the wire guide surface entirely.This will activate the hydrophilic coating.Based on the information provided, inspection of the returned device, and the results of the investigation, the cause for the issue could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute toa death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
| |
|