CARDINAL HEALTH 1.9 FR ARGYLE DUAL LUMEN PICC; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number 43304
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that after the picc dressing was moist, the primary route was observed with medication exiting through the insertion, and the secondary route was observed with good infusion.The catheter was pulled and a hole in the primary route was found.There was a need to remove the catheter.The catheter was not externalized.There was no harm reported.Additional information was received from the customer and stated that the catheter was not out.The catheter was leaking inside the patient`s vein.
 
Additional Manufacturer Narrative
The lot number was provided, and the device history record (dhr) was reviewed indicating that the product was released accomplishing all quality standards.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.A sample was not received for the investigation.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the issue and root cause analysis.At this time, a corrective and preventive action is not deemed necessary.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.9 FR ARGYLE DUAL LUMEN PICC
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
edificio b20 calle #2 zona fra
alajuela
CS  
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key18944480
Report Number3009211636-2024-00804
Device Sequence Number8204186
Product Code LJS
UDI-Device Identifier10884527004652
UDI-Public10884527004652
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number43304
Device Catalogue Number43304
Device Lot Number1922500130
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2024
Supplement Date Received by Manufacturer02/14/2024
Initial Report FDA Received Date03/20/2024
Supplement Report FDA Received Date04/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-