| Catalog Number |
21-7343-24 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: unknown.E1: phone (b)(6).H3: device not received by manufacturer.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the disposable system had become detached from the luer-lock closure without there having been any tension on it.No adverse patient effects were reported by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
D9: date returned to mfg (b)(6) 2024 four samples were received for evaluation.On one sample that was returned in used condition, it was detected that the valve was disconnected from the tubing.On the rest of the samples, no damage or other defects were observed.The reported issue of ¿disconnected¿ was confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
|
| |
|