Catalog Number 07028121190
Medical Device Problem Codes Low Test Results (2458); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
There was an allegation of questionable vitamin b12 g2 elecsys e2g results from the cobas e 801 analytical unit.The initial result was 150 pg/ml with a data flag.The sample was tested on another analyzer and the result was around 270 pg/ml.They repeated the sample on both analyzers and received similar results.No specific data was provided.The customer repeated the sample using a standby pack of the reagent and the result was 263 pg/ml.
 
Additional Manufacturer Narrative
The cobas e 801 analytical unit serial number was (b)(6).The field service engineer checked the analyzer and found no issues.He performed precision testing with acceptable results.The investigation is ongoing.
 
Additional Manufacturer Narrative
Medwatch fields d4 lot no and d4 expiration date were updated.The investigation did not identify a product problem.The specific cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS VITAMIN B12 II
Common Device Name
RADIOASSAY, VITAMIN B12
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18944816
Report Number1823260-2024-00831
Device Sequence Number12857452
Product Code CDD
UDI-Device Identifier04015630939848
UDI-Public04015630939848
Combination Product (Y/N)Y
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K060755
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2024
Device Catalogue Number07028121190
Device Lot Number71148301
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/01/2024
Supplement Date Received by Manufacturer05/23/2024
Initial Report FDA Received Date03/20/2024
Supplement Report FDA Received Date05/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-