| Catalog Number |
21AGFN-756 |
| Medical Device Problem Code |
Material Frayed (1262)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, a 21mm regent aortic mhv,flex c uff was selected for an implant.During the procedure, the leaflet coating peeled from outer frame during implantation.The device was replaced with a 19mm regent aortic mhv,flex c uff that was successfully implanted and completed the procedure.The patient's condition is stable.Abbott employee did not present the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of a leaflet coating peeled from outer frame during implantation was reported.The device was returned to abbott for analysis and the investigation revealed that the bottom of the orifice rim was chipped.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.The cause of the reported incident could not conclusively be determined but is consistent with external force being applied to the valve.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|