| Model Number |
1053-7-50 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The eyeglass tube is coming loose or does not fit at the joint between the tube and the nasal part. .
|
| |
|
Additional Manufacturer Narrative
|
|
The disconnection could prevent the patient from receiving oxygen and become hypoxic.
|
| |
|
Additional Manufacturer Narrative
|
|
The disconnection could prevent the patient from receiving oxygen and become hypoxic.The complaint of " the eyeglass tube is coming loose or does not fit at the joint between the tube and the nasal part." regarding part 1053-7-50 was not confirmed.The root cause was not determined.A risk assessment was performed and the ultimate risk was determined to be low which does not require the initiation of a capa.There have been no other complaints regarding the same part and a similar issue within the 24 months preceding this reported event.A resolution letter was sent to the customer.Complaints will continue to be monitored for possible trends.
|
| |
|
Event or Problem Description
|
|
The eyeglass tube is coming loose or does not fit at the joint between the tube and the nasal part. .
|
| |
|