| Model Number |
EQUINOXE REVERSE 46MM GLENOSPHERE |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
02/21/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section d10: concomitant products - equinoxe reverse tray adapter plate tray +5 (cat# 320-10-05 / serial# (b)(6) ).- eq reverse torque defining screw kit (cat# 320-20-00 / serial# (b)(6) ).- equinoxe reverse 42mm humeral liner +0 (cat# 320-42-00 / serial# (b)(6) ).- eq rev locking screw (cat# 320-15-05 / serial# (b)(6) ).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported, approximately 1.5 years post initial right tsa, the 60 y/o male patient was revised due to dislocation.The patient was revised to a 42mm, +4mm expanded glenosphere and a 42mm liner.The torque screw kit, adapter tray and locking screw were also revised.There was no breakage of device or surgical delay/prolongation patient was last known to be in stable condition following the event.The sales rep was unable to obtain photos/x-rays.The devices are not available for evaluation due to hospital did not return.No other patient information/medical history reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the dislocation reported may have been due to positioning of the components at the time of implantation, mechanical impingement, joint tensioning, or an unreported traumatic event.Based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the dislocation and subsequent revision cannot be conclusively determined.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: corrected medical device problem code.Manufacturer narrative updated: based on review of all available information, there is no evidence to suggest that the reported event is related to any product problems or use issues.The cause of the dislocation cannot be conclusively determined; however, the reported dislocation may have been related to several potential factors including but not limited to the implant selection, the surgical procedure itself, patient anatomical issues, patient activity after surgery and/or underlying patient medical conditions.
|
| |
|