MAUDE Adverse Event Report: MICROVENTION, INC. SCEPTER C; ACCESS DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. SCEPTER C; ACCESS DEVICES Back to Search Results
Model Number BC0410C-ID-700
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/21/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that an onyx embolization treatment in the left ophthalmic artery to treat a tumor.When the embolization procedure was over, the physician tried to remove the balloon.Onyx was adhered to balloon and the balloon catheter did not want to come out.The balloon catheter broke, and some of the balloon catheter remained in the guide catheter and blood vessel.The physician used a trevo device to remove a large segment of the balloon catheter that remained.There was still a fragment of the balloon catheter that remained in the body.The patient was stable.
 
Additional Manufacturer Narrative
Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation, nor were any images of the device provided in place of a device return.Procedure and medical imaging was not provided for this investigation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): potential complications potential complications include, but are not limited to: vessel or aneurysm perforation, vasospasm, hematoma at the site of entry, embolism, ischemia, intracerebral/intracranial hemorrhage, pseudoaneurysm, seizure, stroke, infection, vessel dissection, thrombus formation, and death.Precautions after balloon preparation for use and prior to use, re-inflate to nominal volume and inspect for any irregularities or damage.Do not use if any inconsistencies are observed.The balloon catheter has a lubricious surface and should be hydrated prior to use.Once the balloon catheter is hydrated, do not allow it to dry.Exercise care in handling the balloon catheter to reduce the chance of accidental damage.With the exception of dimethyl sulfoxide (dmso), use of other organic solvents may damage the balloon catheter and/or coating on the surface.Take precaution when manipulating the balloon catheter in tortuous vasculature to avoid damage.Avoid advancement or withdrawal against resistance until the cause of resistance is determined.Presence of calcifications, irregularities or existing devices may damage the balloon catheter and potentially affect its insertion or removal.Excessive torque applied to the syringe might result in damage to the scepter hub assembly.Directions for use (refer to diagram for reference) 1.Attach a rotating hemostatic valve (rhv) to the guidewire lumen of the balloon catheter.Set up a continuous saline flush line and connect it to the sidearm of the rhv.2.Choose appropriate guiding or diagnostic catheter.Attach a rhv to the proximal hub of the guiding or diagnostic catheter.To prevent backflow of blood into the lumen of the catheter, connect the continuous saline flush line to the sidearm of the rhv.3.Open the rhv on the hub of the guiding or diagnostic catheter and introduce the balloon catheter/guidewire into the guiding catheter using the introducer sheath.Carefully advance the balloon catheter/guidewire to the guiding catheter distal tip.After the balloon catheter/guidewire reaches the tip of the guiding catheter, remove the introducer from the balloon catheter shaft by retracting the introducer from the rhv and peeling off the introducer.Advance the balloon catheter through the rhv.4.Advance the balloon catheter and guidewire to the desired location in the vasculature using fluoroscopic visualization.Carefully tighten the valve of the rhv around the balloon catheter to prevent leakage from the rhv.The rhv should still allow for balloon catheter advancement after tightening.Warning: do not over-tighten the rhv around the balloon catheter.Over-tightening could delay balloon inflation and deflation.Warning: do not advance the balloon catheter or guidewire against resistance.If resistance is felt, assess the source of resistance using fluoroscopic means.5.Attach a 2-way stopcock to a 1cc syringe filled with appropriate contrast solution.Prime the 2-way stopcock so that no air is present.Slowly inflate the balloon to the recommended volume to achieve the desired diameter as described in table 3.Warning: do not exceed the maximum recommended inflation volume as balloon rupture may occur.Warning: always inflate and deflate the balloon while visualizing under fluoroscopy to ensure patient safety.6.After inflation, lock the stopcock if desired.7.If desired, remove the guidewire from the balloon catheter and prepare per the respective diagnostic or therapeutic agent ifu(s) for delivery through the guidewire lumen.Warning: excessive pressure higher than 700psi (4826kpa, 47.6atm) may cause leakage or rupture of the guidewire lumen.8.When deflating balloon, use fluoroscopy to ensure complete deflation prior to removal.See table 1 for respective deflation times.After procedure is complete, slowly remove the balloon catheter and guidewire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCEPTER C
Common Device Name
ACCESS DEVICES
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key18946753
Report Number2032493-2024-00232
Device Sequence Number12753942
Product Code MJN
UDI-Device Identifier00842429103388
UDI-Public(01)00842429103388(11)230607(17)260531(10)0000374806
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K110741
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Model NumberBC0410C-ID-700
Device Lot Number0000374806
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age9 YR
Patient SexFemale
-
-