• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN PROTEGE EVERFLEX SELF-EXPANDING STENT SYSTEM STANDARD; STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN PROTEGE EVERFLEX SELF-EXPANDING STENT SYSTEM STANDARD; STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS Back to Search Results
Catalog Number PRB35-07-080-080
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Physician was attempting to use a protege everflex self-expanding stent with a 6fr non-medtronic sheath during a procedure to treat a calcified lesion in the patients right mid superficial femoral artery (sfa).Severe vessel calcification was reported.There were no abnormalities reported in relation to anatomy.There was no damage noted to packaging, i.E.Shelf carton, hoop/tray.There were no issues noted when removing the device from the hoop/tray.The ifu (instruction for use) followed during preparation, procedure, post-procedure without issue.Embolic protection was not used.The vessel was pre and post dilated.The device did not pass through a previously-deployed stent.It was reported the tip of stent catheter broke off after deployment and a snare was used to retrieve.The stent remains in the patient.There was no deformation noted to the deployed stent.A non-medtronic device was used to complete procedure.No patient injury reported.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product analysis.The device was returned with the lock pin loosened, the distal tip of the device appeared to be missing from the device, the device was confirmed from the strain relief, the deployment paddles are approximately 145mm apart the device was unable to be flushed.A 0.035¿ guidewire was unable to advance through the device, it met with resistance, the lumen appeared blocked.The outer blue sheath was skived back with a scalpel and the distal tip was back up in the outer sheath.The distal tip was still attached to the inner sheath, medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGE EVERFLEX SELF-EXPANDING STENT SYSTEM STANDARD
Common Device Name
STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18948621
Report Number2183870-2024-00095
Device Sequence Number7078746
Product Code FGE
UDI-Device Identifier00821684060087
UDI-Public00821684060087
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P110023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/10/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/21/2024
Device Catalogue NumberPRB35-07-080-080
Device Lot NumberB302513
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2024
Supplement Date Received by Manufacturer05/03/2024
Initial Report FDA Received Date03/21/2024
Supplement Report FDA Received Date05/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexFemale
-
-