MEDTRONIC IRELAND TURBOHAWK PLUS 6FR; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Catalog Number THP-S
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Physician was attempting to use a turbohawk plus s directional atherectomy along with a 6fr non-medtronic sheath and a 300cm 0.014 n on-medtronic guidewire during treatment of a fibrous plaque lesion in the patients left mid superficial femoral artery (sfa).Artery diameter reported as 7mm.Lesion exhibited 80% stenosis.There were no abnormalities reported in relation to anatomy.Embolic prote action was not used.There was no damage noted to packaging, i.E.Shelf carton, hoop/tray.There were no issues noted when removing the device from the hoop/tray.The ifu (instruction for use) was followed during preparation, procedure, post-procedure without issue.The vessel was pre-dilated.The vessel was post-dilated with an evercross pta balloon.The device was not passed through a previously deployed stent.No resistance was encountered when advancing the device.It was reported the nosecone separated from drive shaft in the patients vessel.The detached component was removed within the sheath.The device was safely removed from the patient.Procedure was aborted.There was no patient injury/intervention associated with this event.
 
Additional Manufacturer Narrative
Product analysis a visual inspection showed that the tip/housing detached at the hinge pins.The hinge pins are still attached to the device, the reported event that the tip detached can be confirmed based on the condition of the returned device.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURBOHAWK PLUS 6FR
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18948727
Report Number9612164-2024-01373
Device Sequence Number15445153
Product Code MCW
UDI-Device Identifier00763000402419
UDI-Public00763000402419
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K212027
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/20/2026
Device Catalogue NumberTHP-S
Device Lot Number0011918502
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2024
Supplement Date Received by Manufacturer04/19/2024
Initial Report FDA Received Date03/21/2024
Supplement Report FDA Received Date04/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0695-2022
Patient Sequence Number1
Patient SexMale
-
-