| Catalog Number |
119112 |
| Medical Device Problem Codes |
Decrease in Pressure (1490); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the foley catheter spontaneously removed during placement.Per additional information via email from ibc on 08mar2024, it was unknown whether any hole or damage in the balloon.The device used on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter spontaneously removed during placement.Per additional information via email from ibc on 08mar2024, it was unknown whether any hole or damage in the balloon.The device used on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.Three photos were received for evaluation.The first one shows the three-way catheter and pouch.The second photo zooms in the deflated catheter balloon and tip and the last photo shows the catheter cap in which the lot nggp1733 is evidenced.It was also received 1 all silicone temp sensing foley catheter.The catheter balloon was inflated with 5.5 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and no leaks were observed.Balloon concentricity was found within specification.Catheter rested with no leaks.The inhouse syringe was connected and it was able to return the solution with no issues or cuffing in three minutes and 34 seconds.The device is considered within specification, and the reported event is considered unconfirmed.No root cause could be found because the reported event.Dhr is not required since the reported failure was unconfirmed.As the reported event is unconfirmed a labeling review is not required.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|