| Model Number |
8900-0800-01 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to monitor a 50-60-year-old female patient, the electrodes would not adhere to the patient's skin.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Please reference sections d1 brand name updated, d4 catalog # removed, d4 primary udi # updated (this device has a udi; some information was not provided and is unknown; therefore, a complete udi cannot be provided).Evaluation results: zoll medical corporation evaluated the returned pads and retained samples from lot 3023b and the customer's report was not replicated or confirmed.The pads was put through extensive testing including peel testing and visual inspection without duplicating the report.A review of the device history record for lot 2023b showed the raw material passed initial evaluation at incoming inspection.After the evaluation, it was determined that the pads adhere as expected.The returned pads were scrapped after testing.No trend is associated with reports of this type.
|
| |
|
Search Alerts/Recalls
|