| Model Number |
EU2100X15B |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Gas Output Problem (1266); Decrease in Pressure (1490)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3 other text : device not returned to the manufacturer.
|
| |
|
Event or Problem Description
|
|
The manufacturer received information alleging a trilogy evo o2 red alarm and o2 outlet pressure low condition occurred.There was no harm or injury reported.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
Previously reported: the manufacturer received information alleging a trilogy evo o2 red alarm and o2 outlet pressure low condition occurred.There was no harm or injury reported.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.New information: the device was sent to a third-party service center for evaluation.During the evaluation of the device at the third-party service center, error codes were found in the ventilator's downloaded error log.The device's oxygen blending module (obm) and system pca were replaced to address the issue.In addition, the detachable battery and air foam inlet filter were replaced for preventive maintenance.
|
| |
|