| Catalog Number |
ESR916 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
According to the available information the device was explanted and replaced due to a tubing break by the pump.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, non-conforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no non-conforming reports or capa's that were confirmed in veeva to be associated.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan otr pump, cylinders 1 and 2, and reservoir were received for evaluation.Abrasion was noted on both exhaust tubes of the pump.A separation was noted on the longer exhaust tube of the pump, near the strain relief junction.This is a site of leakage.The surfaces appear to have crease marks adjacent to or within a confined area of abrasion, indicating the area was kinked and abrading against itself for an extended period of time.No functional abnormalities were noted with cylinder 1 or cylinder 2.Abrasion was noted on the inlet tube of the reservoir.No functional abnormalities were noted with the reservoir.Based on examination of the returned product, it was concluded that the abrasion marks noted on all pump tubing matched in such a way to conclude that they had overlapped and abraded against one another while in-vivo.This then may have caused the longer exhaust tube of the pump to be kinked onto itself for a period of time.This positioning, in combination with device usage over time, could contribute to sufficient stress to separate the exhaust tubing.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information the device was explanted and replaced due to a tubing break by the pump.
|
| |
|