• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. OPSITE INCISE 28X30CM CTN 10; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. OPSITE INCISE 28X30CM CTN 10; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 4987
Medical Device Problem Codes Delivered as Unsterile Product (1421); Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during set up or inspection for wound treatment, it was noticed that one pouch of a opsite incise 28x30cm ctn 10 had a small hole near the opening of the pouch.The treatment was performed, without any delay, using a s+n back-up dressing.Since incident occurred before procedure; patient was not yet involved.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.All provided information has been reviewed and we have not been able to confirm a relationship between the device and the reported event or determine a definitive root cause.The manufacturing process was reviewed.Dressings are automatically sealed by machine.There are strict inspections during the manufacturing process to ensure that products that do not meet specifications are rejected.Probable cause includes human inspection failure during the packaging process.The batch record was reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of the finished product specification.There was also no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.The complaint history review confirms no additional events of this nature.A review of prior escalation actions found no actions applicable to the scope of this case.This investigation is now complete with no further action deemed necessary.Smith and nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPSITE INCISE 28X30CM CTN 10
Common Device Name
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18950990
Report Number8043484-2024-00023
Device Sequence Number20602535
Product Code KGX
UDI-Device Identifier05000223415501
UDI-Public05000223415501
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number4987
Device Lot Number202328
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2024
Supplement Date Received by Manufacturer05/10/2024
Initial Report FDA Received Date03/21/2024
Supplement Report FDA Received Date05/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-