| Model Number |
3387S-40 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: analysis of the lead (lot #: va2bjuh) revealed that the lead was stretched but no significant anomalies were identified.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported the lead was found to have high resistance and could not be used normally.The lead was replaced with a different lead during surgery, which resolved the issue.The cause of the issue was not determined.The patient was hospitalized for observation after the surgery.
|
| |
|