MPRI ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number 3387S-40
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: analysis of the lead (lot #: va2bjuh) revealed that the lead was stretched but no significant anomalies were identified.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported the lead was found to have high resistance and could not be used normally.The lead was replaced with a different lead during surgery, which resolved the issue.The cause of the issue was not determined.The patient was hospitalized for observation after the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18951072
Report Number2649622-2024-07791
Device Sequence Number12525410
Product Code MHY
UDI-Device Identifier00643169752290
UDI-Public00643169752290
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P960009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/12/2023
Device Model Number3387S-40
Device Catalogue Number3387S-40
Device Lot NumberVA2BJUH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/18/2021
Initial Report FDA Received Date03/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-