HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSISST) BYPASS
|
Back to Search Results |
|
| Model Number |
1650 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Continuation of d10: ddpa2d4 icd implanted (b)(6) 2021 407652 lead implanted (b)(6) 2014 6935m6 lead implanted (b)(6) 2014, additional products: d1: heartware ventricular assist system ¿ battery d4: model#: 1650 / catalog#: 1650 / expiration date: 31-jan-2023 /serial#: (b)(6), udi#: (b)(4), d9: no h3: no h4: mfg date: 24-jan-2022 h5: no h6: the codes present in section h6 correspond to components/products that comprise the reported event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that after routine log file review, the batteries exhibited communication errors.The batteries remain in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report is being submitted for investigation summary.Product event summary: two (2) batteries (bat (b)(6) , bat (b)(6) ) were not returned for evaluation.Log file analysis revealed that the controller in use during the reported event, (b)(6), contained a feature that records whether a power source experienced a communication error or a disconnection within each 15-minute interval.Review of the data log file revealed instances involving bat (b)(6) and bat (b)(6) where the relative state of charge (rsoc) value was logged between 101-201, which is indicative of a communication error.As a result, the reported event was confirmed.The batteries were lubricated prior to release.Possible root causes of the reported communication errors can be attributed to a momentary disconnection on the communication pins of the controller, the controller not receiving responses from the batteries, and/or due to the packet error checking method detecting bit errors.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|