MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD IAG BC PRO GLOBAL WING; PERIPHERAL IV CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD IAG BC PRO GLOBAL WING; PERIPHERAL IV CATHETERS Back to Search Results
Catalog Number 382933
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.B3.The date received by manufacturer has been used for this field.
 
Event or Problem Description
It was reported that bd iag bc pro global wing package seal integrity.The following information was provided by the initial reporter, translated from french to english: i am sending you a material safety report of zero clinical importance.We found 11 units of unsealed (primary packaging unsealed) and therefore non-sterile.We have not found this defect in other boxes from the same batch.We're not asking for credit or free units, i just wanted to let you know.
 
Event or Problem Description
No additional information.
 
Additional Manufacturer Narrative
Investigation results: the complaint of unsealed packaging was confirmed and the cause appeared to be packaging related.Two photographs and eleven 20ga insyte autoguard units from lot #2340957 were provided for investigation.The eleven units did not exhibit a complete seal due to the top web being misaligned with the bottom web.This defect likely occurred during the packaging process.During the sealing process, it appeared that the top web was misaligned with the bottom web, which resulted in an incomplete and open seal.100% vision system and setup verification per procedures are in place to mitigate the occurrence of this defect.The appropriate manufacturing personnel were notified of this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD IAG BC PRO GLOBAL WING
Common Device Name
PERIPHERAL IV CATHETERS
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18951312
Report Number1710034-2024-00211
Device Sequence Number15430268
Product Code FOZ
UDI-Device Identifier00382903829330
UDI-Public(01)00382903829330
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
UNK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/30/2025
Device Catalogue Number382933
Device Lot Number2340957
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2024
Supplement Date Received by Manufacturer06/05/2024
Initial Report FDA Received Date03/21/2024
Supplement Report FDA Received Date06/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-