| Model Number |
M00534360 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Premature Activation (1484); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an advanix biliary stent with naviflex rx delivery system was to be implanted during a procedure on (b)(6) 2023.During the procedure, the stent separated from the holder by itself and got stuck in the endoscope.There were no patient complications reported as a result of this event.Note: this event has been deemed an mdr-reportable event based on the investigation finding of stent damaged.Please see block h10 for the full investigation details.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: the initial reporter's facility name is (b)(6).The initial reporter's address is 29 av du mal delattre de tassigny.H6: imdrf device code a0406 captures the reportable investigation finding of stent damage.H10: the advanix biliary stent and naviflex rx delivery system were received for analysis.Visual inspection revealed that the stent was damaged.No other damages were noted with the stent or the delivery system.Product analysis was unable to confirm the reported events of stent premature deployment and device or device component entrapment because these events occurred during the procedure and were not possible to replicate in the laboratory of analysis.However, the investigation concluded that the additional investigation finding of a damaged stent was most likely due to procedural factors such as lesion characteristics, the handling of the device, or the technique used by the physician (force applied).Taking all available information into consideration, the root cause of the reported clinical observation is no problem detected.A product labeling review identified that the device was used per the instructions for use (ifu)/product label.
|
| |
|
Search Alerts/Recalls
|