• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTERLOCK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTERLOCK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83779
Medical Device Problem Codes Premature Activation (1484); Failure to Advance (2524); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the coil was removed with a snare.The target lesion was located in the severely tortuous inferior mesenteric artery.After placing a renegade stc-18 microcatheter, a 6mm x 20cm interlock was selected for use.However, during the procedure, the coil was not pushed smoothly, and the catheter popped out after the obstruction resulting in the detachment of the coil inside the patient.The coil was removed with a snare.A 35-system, 8mm x 40cm interlock with a 5f non-boston scientific angiographic catheter were selected for embolization and reduction.The angiography was normal, and the procedure was completed successfully.There were no complications reported and the patient was stable post procedure.
 
Event or Problem Description
It was reported that the coil was removed with a snare.The target lesion was located in the severely tortuous inferior mesenteric artery.After placing a renegade stc-18 microcatheter, a 6mm x 20cm interlock was selected for use.However, during the procedure, the coil was not pushed smoothly, and the catheter popped out after the obstruction resulting in the detachment of the coil inside the patient.The coil was removed with a snare.A 35-system, 8mm x 40cm interlock with a 5f non-boston scientific angiographic catheter were selected for embolization and reduction.The angiography was normal, and the procedure was completed successfully.There were no complications reported and the patient was stable post procedure.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, the main coil, introducer sheath and the pusher wire were returned.It was observed that the main coil was bent and stretched at all primary coil section.The functional inspection could not be performed due the main coil and the pusher wire are not interlocking.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18952143
Report Number2124215-2024-16539
Device Sequence Number15485303
Product Code KRD
UDI-Device Identifier08714729765004
UDI-Public08714729765004
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/08/2026
Device Model Number83779
Device Catalogue Number83779
Device Lot Number0031778553
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/06/2024
Supplement Date Received by Manufacturer05/03/2024
Initial Report FDA Received Date03/21/2024
Supplement Report FDA Received Date05/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/08/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROCATHETER: RENEGADE STC-18; MICROCATHETER: RENEGADE STC-18
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexFemale
-
-