| Model Number |
G5A-80A |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
Necrosis (1971)
|
| Date of Event |
02/28/2024
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing without duplicating the report.The device was recertified and returned to the customer.Review of the data log showed the device properly identified pads connected, and no pads-related or device errors were detected other than user advisories to check pad connection to the patient.No trend is associated with reports of this type.An internal inspection resulted in no findings.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a female patient (age unknown), the device displayed a "check pads" message.Complainant indicated that the clinician obtained another set of defib pads to continue treating the patient, however, the same error was observed.The complainant indicated that the ems team arrived shortly after and their equipment was used to treat the patient with no success.Complainant indicated that the patient subsequently expired.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|