ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1804250-23
Medical Device Problem Codes Difficult to Advance (2920); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that during the procedure an unspecified guide wire (gw) was in the left anterior descending (lad) coronary and another unspecified gw was in the left circumflex (lcx) coronary artery with stenting intended for the lcx.The 2.5x23 xience skypoint stent delivery system (sds) felt resistance advancing over the gw.The stent was noted to move on the balloon and was removed from the anatomy intact.Another xience skypoint sds was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
A visual, functional, and dimensional inspection was performed on the returned device.The reported stent dislodgement and difficult to advance/position was not confirmed.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database identified no similar complaints from this lot.A conclusive cause for the reported difficulties could not be determined; however, stent dislodgement may be attributed to several factors including, but not limited to, improper or inadequate crimping at the time of manufacture, incorrect sheath sizing, positive pressure during preparation, forced sheath removal, handling of the stent during preparation, interaction with the anatomy, or interaction with accessory devices.Difficult to advance/position may be attributed to several factors including, but are not limited to, product placement technique, guiding catheter support, inner diameter of guide wire lumen, outer diameter of the guide wire, patient morphology, condition of the guide wire, interaction with accessory devices, damage to the catheter or accumulation of blood or contrast.In this case, it is possible that the device may have interacted with accessory devices (two guide wires inserted) during advancement, as resistance was noted, contributing to the reported difficult to advance/position.Further interaction with accessory devices during advancement may have contributed to the observed material deformation (distal stent damage) which the account initially reported as a stent dislodgement; however, this cannot be confirmed.There is no indication of a product quality issue with respect to manufacture, design, or labeling; therefore, no product-related corrective action will be implemented in this case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18953795
Report Number2024168-2024-03595
Device Sequence Number18182868
Product Code NIQ
UDI-Device Identifier08717648233098
UDI-Public(01)08717648233098(17)260917(10)3092041
Combination Product (Y/N)Y
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/17/2026
Device Catalogue Number1804250-23
Device Lot Number3092041
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/04/2024
Supplement Date Received by Manufacturer06/12/2024
Initial Report FDA Received Date03/21/2024
Supplement Report FDA Received Date06/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexMale