Catalog Number 125040
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the device did not inflate post insertion in the theatre".As a result, the patient was extubated and a new lma unique was inserted.No patient harm or injury.The patient status is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.Other remarks: n/a.Corrected data: n/a.
 
Event or Problem Description
It was reported that "the device did not inflate post insertion in the theatre".As a result, the patient was extubated and a new lma unique was inserted.No patient harm or injury.The patient status is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).One unit was received for evaluation.Sample size was the sample as complaint description.From the visual inspection no abnormality found.Leak test was performed to actual complaint sample to review the functionality of the device.Cuff been soaked into the water and no presence of bubble at cuff area & jointing found.By deflating the cuff fully and inflate the cuff as per the instruction for use.No abnormality found, cuff can be inflated and deflated normally.This step was executed 10 times.A device history record review was performed, and no relevant findings were identified.Since functional test already been performed to the actual complaint sample and no abnormality found, complaint concluded as no problem found on sample.This complaint identifies as non-manufacturing related since no problem found on sample.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that "the device did not inflate post insertion in the theatre".As a result, the patient was extubated and a new lma unique was inserted.No patient harm or injury.The patient status is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LMA UNIQUE SIZE 4
Common Device Name
LMA UNIQUE
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
THE LARYNGEAL MASK COMPANY (M) SDN. BHD
lot 19 , jalan hi-tech 3
zon industri fasa 1, kulim hi-tech park
kulim kedah 09090
MY   09090
Manufacturer Contact
mariah mackinnon
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18953820
Report Number3009307931-2024-00012
Device Sequence Number12532251
Product Code CAE
UDI-Device Identifier15060112314718
UDI-Public15060112314718
Combination Product (Y/N)N
Initial Reporter CountrySF
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/02/2025
Device Catalogue Number125040
Device Lot Number11F22G0033
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2024
Supplement Date Received by Manufacturer04/05/2024
05/20/2024
Initial Report FDA Received Date03/21/2024
Supplement Report FDA Received Date04/23/2024
05/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age34 YR
Patient SexFemale
-
-