Visual inspections were performed on the returned device.The reported failure to fire was not confirmed.Unused sterile devices were returned from the account.The sterile units were visually and functionally tested with no anomalies noted.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no similar incidents from this lot.Factors that may contribute to the reported failure to deploy the needles may include, but are not limited to, change position of device during deployment, device orientation, over insertion of the device.The reported difficulties and treatment appears to be related to circumstances of the procedure.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|