• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5A
Medical Device Problem Codes Deflation Problem (1149); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported, the flexible scope exhibited a leak.The issue occurred during a therapeutic stent replacement.There were no reports of patient harm.
 
Event or Problem Description
It was reported that there was damage to the tube mounting bracket [port].The issue was found during reprocessing.There was no report of patient impact.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: b3, b5, h6 component code, h6 problem code.New information added to the following field: h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over eight (8) years since the subject device was manufactured.The device evaluation confirmed the reported event and found that the instrument channel/tube mounting port was damaged.Based on the results of the investigation, a definitive root cause could not be determined.The event can be prevented by following the instructions for use which state: "oes cystonephrofiberscope olympus cyf-5a important information ¿ please read before use warnings and cautions ¿ do not apply shock to the distal end of the endoscope.Visual abnormalities may result.¿ do not twist or bend the bending section with your hands.Equipment damage may result.¿ do not squeeze the bending section forcefully.The covering of the bending section may stretch or break and cause water leaks." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Common Device Name
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18955442
Report Number3002808148-2024-02730
Device Sequence Number11907168
Product Code FAJ
UDI-Device Identifier04953170339424
UDI-Public04953170339424
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K032092
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCYF-5A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer04/08/2024
Initial Report FDA Received Date03/21/2024
Supplement Report FDA Received Date04/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-