| Model Number |
12322-01 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that an arteriotomy closure of the common femoral artery was attempted with three prostar devices related to an interventional procedure with a 10f sheath.Reportedly, the needle did not line up with the opening for all three devices.An additional prostar device was used to achieve hemostasis.There was no adverse patient sequela and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.The two additional prostar devices referenced in b5 are filed under separate medwatch report numbers.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspections were performed on the returned device.The reported failure to fire was not confirmed.Unused sterile devices were returned from the account.The sterile units were visually and functionally tested with no anomalies noted.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history did not indicate a lot specific quality issue.Factors that may contribute to the reported failure to deploy the needles may include, but are not limited to, change position of device during deployment, device orientation, over insertion of the device.The reported difficulties and treatment appears to be related to circumstances of the procedure.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|