EXACTECH, INC. NV GXL LNR, 10 DEG FACE, 32MM ID, GRP 1 CUP; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
138-32-51 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
11/02/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported via legal documentation that a patient had a right total hip arthroplasty on (b)(6) 2015, and then experienced a revision surgical procedure on (b)(6)(2023, approximately 8 years, 5 months after initial implant.There was no other patient/medical information provided.No x-rays or images were provided.The device will not be returned.There is no other information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending investigation.There is no other information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The most likely cause for the reported revision due to prosthesis wear in is a combination of the risk factors: such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|