MEDOS INTERNATIONAL SARL OG TDL CMPLX FILL COIL 5X15; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Catalog Number 640CF0515
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Section d.2b: procode is krd/hcg.Section e.1: the initial reporter phone and email address are not available / reported.Section h.3: the device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.A review of manufacturing documentation associated with this lot (30842966) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
The healthcare professional reported that during an endovascular embolization procedure, the 5mm x 15cm orbit galaxy complex fill coil (640cf0515 / (b)(6)) got detached inside the concomitant microcatheter (unspecified brand) during its deployment inside the aneurysm.The procedure was delayed by 5 minutes.The complaint coil was removed.There was no negative patient impact.The procedure was reported as successfully completed.On 12-mar-2024, additional information was received.Per the information, the target was a 5.9mm x 5.2mm internal carotid artery (ica) supraclinoid aneurysm.The information indicated that the coil did not prematurely detach at the detachment zone on the device positioning unit; the coil did not become separated into two or more pieces.The microcatheter used was not a cerenovus microcatheter.The same microcatheter was used with the replacement coil, a 5mm x 10cm orbit galaxy complex fill coil ((b)(6)).The physician did not consider the 5-minute extension in the procedure to be clinically significant.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No:(b)(4).The purpose of this mdr submission is to report the investigational finding of the returned device.The complaint product was returned and received for evaluation and analysis.The investigation is documented below.Investigation summary: a non-sterile 5mm x 15cm orbit galaxy complex fill coil was received contained in the decontamination pouch.Visual inspection was performed.There was no appearance of damage observed.The returned device was inspected under the microscope.Under magnification, it was observed that the embolic coil was no longer attached to the gripper and this was not returned for evaluation.No elongation marks were found on the gripper.The issue regarding the premature detachment of the coil was confirmed since the coil was no longer attached to the returned gripper component.However, no damages were found on the device which indicates the detachment occurred prematurely as a result of a device failure.According to the risk documentation, premature detachment of the embolic coil assembly from the delivery system is a potential failure that can occur during embolic coil placement due to manipulation of the system against resistance.No further evaluation of contributing factors can be conducted without the embolic coil.It is possible that other clinical and procedural factors not described in this complaint may have contributed to the reported failure.A review of manufacturing documentation associated with this lot (30842966) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) contains the following precaution: ¿ if unusual friction is still noticed during advancement or retraction of the microcoil system, verify flush lines are open and properly pressurized.Then slowly withdraw the entire microcoil system and examine for damage.Replace it with a new microcoil system.If friction still exists, withdraw and examine the delivery catheter system.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that the product analysis lab received the complaint device on 10-apr-2024.A supplemental 3500a report will be submitted once the product investigation has been completed.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OG TDL CMPLX FILL COIL 5X15
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvarcar
juarez chihuahua 32574
MX   32574
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
303552-6892
MDR Report Key18956522
Report Number3008114965-2024-00282
Device Sequence Number9445597
Product Code KRD
UDI-Device Identifier10886704030300
UDI-Public10886704030300
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K093973
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Catalogue Number640CF0515
Device Lot Number30842966
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2024
Supplement Date Received by Manufacturer04/10/2024
04/15/2024
Initial Report FDA Received Date03/21/2024
Supplement Report FDA Received Date04/10/2024
04/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROCATHETER (UNSPECIFIED BRAND)
Patient SexUnknown
-
-