MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. THUNDERBEAT 5 MM, 35 CM, PISTOL GRIP; ULTRASONIC SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. THUNDERBEAT 5 MM, 35 CM, PISTOL GRIP; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0535PC
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported during a therapeutic hysterectomy the probe fell off into the patients abdominal cavity.The probe was retrieved with another instrument and there was no effect to the patient's health.The procedure was completed with a similar device.
 
Additional Manufacturer Narrative
This report is being submitted to provide the legal manufacturer's final investigation results.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the probe likely encountered other surgical instruments, or non-insulated areas as described below in the mechanisms below: mechanism #1 1.Hard tissue, a metal object or a surgical instrument had been in contact with the probe during the output activation in seal & cut mode causing scratches on the probe.2.The device was activated in seal & cut mode or while the grasping section was grasping tissue.This applied force to the scratched area of the grasping section, causing this area to crack.3.A force was applied to the probe causing it to break.Mechanism #2 1.Grasping section was closed without grasping anything while the device was activating in seal & cut mode (this includes after tissue resection) causing the tissue pad to wear out.2.Since the tissue pad was worn out, the non-insulated area of the grasping section and the distal end of the probe came into contact.3.The output was activated in seal & cut mode in the state of description stated above.Therefore, scratches (contact marks) were made on the distal end of the probe and the grasping section, indicating that they encountered each other.4.The device was activated in seal &cut mode or while the grasping section was grasping tissue.This applied force to the scratched area of the grasping section, causing this area to crack.5.A force was applied to the probe causing it to break.However, the root cause of the reported event could not be determined.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ ¿the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator, forceps, and others).Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/splitting/protruding/partial separating of the tissue pad.In turn, the probe may break before displaying an error window or generating an alarm tone.¿ ¿do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.¿ this supplemental report includes a correction to g2 to provide information that was inadvertently not included in the initial medwatch.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 35 CM, PISTOL GRIP
Common Device Name
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18956609
Report Number9614641-2024-00750
Device Sequence Number13097967
Product Code GEI
UDI-Device Identifier04953170308659
UDI-Public04953170308659
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K132703
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTB-0535PC
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2024
Supplement Date Received by Manufacturer04/22/2024
Initial Report FDA Received Date03/21/2024
Supplement Report FDA Received Date04/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-