MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SYNCHRONY; COCHLEAR IMPLANT Back to Search Results
Model Number MI1200 SYNCHRONY PIN
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Necrosis (1971)
Date of Event 03/19/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The recipient experienced skin necrosis over the implant magnet and previously had skin irritation with the magnet in the last several months.Reportedly the implant is working without failure.The explantation surgery is scheduled for (b)(6) 2024.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Additional Manufacturer Narrative
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.According to the information received from the field the device was explanted due to a skin necrosis above the implant.A review of the device_s sterilization records shows that the device has been subject to a valid sterilization process, thus no available information points to the implant being the source of reported issue.However, the presence of the device might have contributed to its subsequent development.
 
Event or Problem Description
The recipient experienced skin necrosis over the implant magnet and previously had skin irritation with the magnet in the last several months.Reportedly the implant is working without failure.The user was explanted on (b)(6) 2024.
 
Additional Manufacturer Narrative
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.According to the information received from the field the device was explanted due to a skin necrosis above the implant.A review of the device's sterilization records shows that the device has been subject to a valid sterilization process, thus no available information points to the implant being the source of reported issue.However, the presence of the device might have contributed to its subsequent development.This is a final report.
 
Event or Problem Description
The recipient experienced skin necrosis over the implant magnet and previously had skin irritation with the magnet in the last several months.Reportedly the implant is working without failure.The user was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key18957623
Report Number9710014-2024-00274
Device Sequence Number14572504
Product Code MCM
UDI-Device Identifier09008737094034
UDI-Public(01)09008737094034
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/27/2022
Device Model NumberMI1200 SYNCHRONY PIN
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2024
Supplement Date Received by Manufacturer03/19/2024
03/19/2024
Initial Report FDA Received Date03/22/2024
Supplement Report FDA Received Date06/21/2024
11/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/27/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age7 YR
Patient SexFemale
-
-