1038671-2024-00639 b1: corrected.B2: corrected.C6: corrected the following: health effect - clinical code, type of investigation, investigation findings, investigation conclusions.Manufacturer narrative: the most likely cause for the reported revision due to prosthesis wear and osteolysis is a combination of the risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).Additionally, a contributing factor may have been using another manufacturer¿s femoral components with exactech acetabular components.However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.
|