Event or Problem Description
|
|
It was reported that the implant abutment alongside the restoration was separated from the implant and an abutment fracture was discovered.X-rays also confirmed that a piece of abutment was fractured inside the implant.The implant was removed.Tooth site #26.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint (b)(4).A4: patient weight is not provided / unknown.D1: brand name is not provided / unknown.D4: catalog number and lot number are not provided / unknown.D10: tsv4b10, imp,tsv,4.1mm,sbm,10, lot number 63847812.E1: initial reporter¿s title is not provided / unknown.
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to report additional information.Follow up with customer confirmed item and lot number.G4: additional 510(k) numbers are k013227 and k953101.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to report additional information.Zimvie did not receive one abutment for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was patient factors.Therefore, based on the available information, a device malfunction could not be verified without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|