• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL ABUT HEX-LOCK 4.5MM IMP 5 .5 FLARE; DENTAL ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL ABUT HEX-LOCK 4.5MM IMP 5 .5 FLARE; DENTAL ABUTMENT Back to Search Results
Catalog Number HLA4/5
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/11/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant abutment alongside the restoration was separated from the implant and an abutment fracture was discovered.X-rays also confirmed that a piece of abutment was fractured inside the implant.The implant was removed.Tooth site #26.
 
Additional Manufacturer Narrative
Zimvie complaint (b)(4).A4: patient weight is not provided / unknown.D1: brand name is not provided / unknown.D4: catalog number and lot number are not provided / unknown.D10: tsv4b10, imp,tsv,4.1mm,sbm,10, lot number 63847812.E1: initial reporter¿s title is not provided / unknown.
 
Event or Problem Description
No additional or corrected information to report.
 
Additional Manufacturer Narrative
This report is being submitted to report additional information.Follow up with customer confirmed item and lot number.G4: additional 510(k) numbers are k013227 and k953101.
 
Additional Manufacturer Narrative
This report is being submitted to report additional information.Zimvie did not receive one abutment for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was patient factors.Therefore, based on the available information, a device malfunction could not be verified without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional or corrected information to report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABUT HEX-LOCK 4.5MM IMP 5 .5 FLARE
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key18957760
Report Number0002023141-2024-00891
Device Sequence Number18503192
Product Code NHA
UDI-Device Identifier00889024015210
UDI-Public(01)00889024015210(17)261211(10)2021120242
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K011028
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberHLA4/5
Device Lot Number2021120242
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date03/22/2024
Supplement Report FDA Received Date04/04/2024
08/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/10/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DENTAL IMPLANT, SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-