MAUDE Adverse Event Report: MODERN MEDICAL EQUIP MFG LTD PLUMEPEN ELITE SURGICAL SMOKE EVAC PENCIL, 15FT TUBING,QTY20; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MODERN MEDICAL EQUIP MFG LTD PLUMEPEN ELITE SURGICAL SMOKE EVAC PENCIL, 15FT TUBING,QTY20; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number PLP2520
Medical Device Problem Code Self-Activation or Keying (1557)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the plp2520, plumepen elite surgical smoke evac pencil, 15ft tubing,qty20 was being used om (b)(6) 2024 during a lumbar laminectomy and "the bipolar electrocautery ws not working.They changed the bipolar cord and noticed that it still wasn't working.After further investigation it was recognized that the cautery was still activated.Bipolar will not work if cautery is activated.Cautery pencil was in the protective holster and not to be used at this time.Cautery pencil was removed from the filed.A new pencil was given.At that time the bipolar started working.¿.There was no impact or injury to the patient.The procedure was completed.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
The device is not expected to be returned for evaluation and review.However, the complaint investigation is not complete at this time.A supplemental and final report will be filed following the completion of the complaint investigation.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Event or Problem Description
The customer reported that the plp2520, plumepen elite surgical smoke evac pencil, 15ft tubing,qty20 was being used om 1mar24 during a lumbar laminectomy and "the bipolar electrocautery ws not working.They changed the bipolar cord and noticed that it still wasn't working.After further investigation it was recognized that the cautery was still activated.Bipolar will not work if cautery is activated.Cautery pencil was in the protective holster and not to be used at this time.Cautery pencil was removed from the filed.A new pencil was given.At that time the bipolar started working.¿.There was no impact or injury to the patient.The procedure was completed.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
The device will not be returned and no photographic evidence was provided therefore, a device malfunction cannot be verified.A 2 year lot history review shows a total of 5 devices for this lot number and failure mode however, none are confirmed.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.(b)(4) we will continue to monitor for trends through the complaint system to assure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUMEPEN ELITE SURGICAL SMOKE EVAC PENCIL, 15FT TUBING,QTY20
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MODERN MEDICAL EQUIP MFG LTD
unit a, 10/f mai wah ind bldg
1-7 wah sing street
kwai chung, nt
HK 
Manufacturer (Section G)
MODERN MEDICAL EQUIP MFG LTD
unit a, 10/f mai wah ind bldg
1-7 wah sing street
kwai chung, nt
HK  
Manufacturer Contact
robin drum
11311 concept blvd
largo, FL 33773
8653881978
MDR Report Key18957929
Report Number1320894-2024-00077
Device Sequence Number4591625
Product Code GEI
UDI-Device Identifier10845854061305
UDI-Public(01)10845854061305(17)250626(10)MM20220688
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K230547
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/26/2025
Device Catalogue NumberPLP2520
Device Lot NumberMM20220688
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2024
Supplement Date Received by Manufacturer03/26/2024
Initial Report FDA Received Date03/22/2024
Supplement Report FDA Received Date04/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-