| Model Number |
IMMULITE 2000 XPI SYSTEM |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Falsely depressed insulin-like growth factor binding protein 3 (igfb-3) results were obtained on multiple patient samples on an immulite 2000 xpi instrument.The erroneous results were not reported to the physician(s).The samples were repeated on an alternate immulite 2000 xpi instrument.The repeat results were higher than the erroneous results.The repeat results were reported, as the correct results, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the falsely depressed insulin-like growth factor binding protein 3 (igfb-3) results.
|
| |
|
Additional Manufacturer Narrative
|
|
An outside the us customer contacted a siemens customer care center to report falsely depressed insulin-like growth factor binding protein 3 (igfb-3) results on multiple patient samples on an immulite 2000 xpi instrument.Siemens is investigating.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens filed the initial mdr 2247117-2024-00002 on 22-march-2024.Additional information 21-june-2024: siemens evaluated this issue and provided the following conclusion: the customer performed a 10-patient comparison study on the complaint analyzer and an alternate immulite 2000 xpi.The data was analyzed and showed valid adjustments and quality control recovering within acceptable ranges.In addition, all patient samples recovered within 10% instrument-to-instrument.The customer is operational.The immulite 2000 xpi insulin-like growth factor binding protein 3 (igfbp-3) assay, lot 431, and the immulite 2000 xpi system are performing as intended.A product performance issue has not been identified.No further evaluation of the device is required.In section h6, the investigation findings and investigation conclusions codes were updated.
|
| |
|
Search Alerts/Recalls
|