C.R. BARD, INC. (COVINGTON) -1018233 BARD® TOUCHLESS® PLUS UNISEX PRE-LUBRICATED VINYL URETHRAL CATHETER Back to Search Results
Catalog Number 4A6144
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that intermittent catheters did not have holes in it.Per follow up via phone on (b)(6) 2024, it was reported that customer kept received catheters with no holes and they need them to be replaced.Customer purchased the catheters from bard.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was available for evaluation and further investigation did not result in any additional findings.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be ¿omission of operation".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labeling could not have prevented the reported failure.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that intermittent catheters did not have holes in it.Per follow up via phone on 13mar2024, it was reported that customer kept received catheters with no holes and they need them to be replaced.Customer purchased the catheters from bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® TOUCHLESS® PLUS UNISEX PRE-LUBRICATED VINYL URETHRAL CATHETER
Common Device Name
PRE-LUBRICATED VINYL URETHRAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18958192
Report Number1018233-2024-01508
Device Sequence Number9290889
Product Code FCM
UDI-Device Identifier00801741030604
UDI-Public(01)00801741030604
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K910653
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number4A6144
Device Lot NumberNGHW2877
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2024
Supplement Date Received by Manufacturer04/18/2024
Initial Report FDA Received Date03/22/2024
Supplement Report FDA Received Date04/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-