| Catalog Number |
300613 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Seal on the plunger has a piece of rubber on it.Additional information from customer: the piece of rubber is attached to the stopper.
|
| |
|
Additional Manufacturer Narrative
|
|
Pr (b)(4): initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.H3 other text : see h10 manufacture narrative.
|
| |
|
Additional Manufacturer Narrative
|
|
Pr (b)(4) follow up mdr for device evaluation: two photos and one physical sample were provided to our quality team for investigation.Through visual inspection, a black piece which corresponds to the stopper, was observed within the syringe.A device history review was performed for the reported lot 2312046, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this incident.The stopper is provided by an external supplier.Incoming inspections are performed according to procedure, including verification the stoppers are complete and not deformed.The quality certificate was reviewed for the stopper batches used with syringes from lot 2312046, no issues were identified.Retained samples of the reported lot were used for additional evaluation.All stoppers were inspected and no damage or other defects were observed on any of the devices.Final products in this manufacturing line, for this reference are sampled and they are subjected to visual and functional inspections during the different manufacturing sub-processes according to procedures, no issues related to the reported incident were found.Based on the sample evaluation, it is believed this incident is related to an issue during the stopper manufacturing.We have notified the supplier of this incident.Complaints received for this device and reported condition will be monitored by our quality team for signs of emerging trends.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|