| Catalog Number |
363LF-690 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/31/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "deflated implant and textured saline device." this record is for the right side.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflated implant and textured saline device." this record is for the right side.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: deflation: observed opening device assessed as surgical damage.Anxiety-product-procedure: unable to observe since it is not related to the device.No further actions are required as no manufacturing issues are observed.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflated implant and textured saline device." this record is for the right side.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to g.3.Aware date of supplemental medwatch #1.Aware date should have been listed as 23/apr/2024.
|
| |
|