| Model Number |
BD-410X-1855 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, the ezdilate balloon dilator (wg) 16-17-18 was filled up the first time and it fulfilled its function.When the balloon was refilled, it was leaking.The issue occurred during an endoscopic retrograde cholangiopacification (ercp) procedure.The procedure was completed using a similar device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The customer stated that the device was test inflated with air before it was inflated again for the procedure.The device is single-use and not designed to be inflated more than once.This can be found in the instructions for use.The device was not returned to olympus for evaluation.Olympus will continue to monitor field performance for this device.
|
| |
|