| Catalog Number |
04810716190 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter received a questionable crep2 (creatinine plus ver.2) result from one patient sample tested on the cobas 4000 c311 stand alone system.The initial result was reported outside of the laboratory; a complaint was received prompting the rerun of the patient sample.On (b)(6) 2024: the initial result was 6.32 mg/dl.The first repeat result was 0.59 mg/dl.On (b)(6) 2024: the second repeat result was 0.59 mg/dl.The repeat results were consistent with the patient's clinical condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The serial number of the customer's cobas 4000 c311 stand alone system is (b)(6).The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The field service representative (fsr) investigated the event and the customer's processing of patient samples.The fsr determined the event was consistent with the pre-analytical processing of the patient sample.The investigation determined that based on the fsr's investigation, the event was related to a preanalytic issue at the customer site.Preanalytics are within the customer's responsibility.There was no indication of a device malfunction.
|
| |
|