| Model Number |
8881450004 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported that during oral suctioning, the plastic part on the edge came off.The patient swallowed the fallen part accidentally, and, eventually it was retrieved with an endoscope.There was no additional medication or no further medical treatment.The patient appeared to be advanced age since he/she has no teeth at all.The period of use was a couple of minutes.There was no infection.It did fall/retain in the body.There is no further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample was received at the plant for the investigation.Because a lot number was not provided, the device history record could not be performed.Upon a visual evaluation of the sample, the reported issue was confirmed; detachment was observed between the tube and connector.A root cause was unable to be determined.Actions have been taken to address the reported condition.A quality alert was also issued to alert manufacturing personnel as part of the quality containment.
|
| |
|