• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION POLARIS ULTRA; STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION POLARIS ULTRA; STENT, URETERAL Back to Search Results
Model Number M0061921230
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: initial reporter facility name: (b)(6) capital medical university.H6: imdrf device code a0401 captures the reportable event of stent shaft broken.
 
Event or Problem Description
It was reported to boston scientific corporation that a polaris ultra ureteral stent was used during a ureteral catheterization procedure in the ureter performed on (b)(6) 2021.During the procedure, it was noted that the front-end part of the stent was found fractured.The procedure was successfully completed with another polaris ultra ureteral stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.A photo of the complaint device was provided and showed that the stent shaft was broken.
 
Event or Problem Description
It was reported to boston scientific corporation that a polaris ultra ureteral stent was used during a ureteral catheterization procedure in the ureter performed on (b)(6) 2021.During the procedure, it was noted that the front-end part of the stent was found fractured.The procedure was successfully completed with another polaris ultra ureteral stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.A photo of the complaint device was provided and showed that the stent shaft was broken.
 
Additional Manufacturer Narrative
Block e1: initial reporter facility name: (b)(6).Block h6: imdrf device code a0401 captures the reportable event of stent shaft broken.Block h10: the returned polaris ultra ureteral stent was analyzed, and a visual and microscopic evaluation noted that the shaft detached.Additionally, the suture and positioner did not return.Media inspection showed the shaft detached and the box of the device, additionally, the information of the box matches with the information of the complaint.No other problems with the device were noted.The reported event of was confirmed.Taking all available information into consideration, most likely, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.It is possible to conclude that this problem could be caused by operational factors.The retrieval line may be removed before placement at the physician's discretion.If the retrieval line gets entangled in the shaft and is removed using excess force, the stent can get detached during the procedure affecting the performance of the device.Therefore, all compiled information on this investigation determines that the most probable cause is "adverse event related to procedure" since the adverse event occurred during the procedure and the device had no influence on event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARIS ULTRA
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18962020
Report Number2124215-2024-15594
Device Sequence Number8052443
Product Code FAD
UDI-Device Identifier08714729124320
UDI-Public08714729124320
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K010002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/09/2026
Device Model NumberM0061921230
Device Catalogue Number192-123
Device Lot Number0031209361
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/26/2024
Supplement Date Received by Manufacturer04/11/2024
Initial Report FDA Received Date03/22/2024
Supplement Report FDA Received Date05/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age53 YR
Patient SexMale
Patient Weight61 KG
-
-