MAUDE Adverse Event Report: TORNIER S.A.S. TORNIER FLEX SHOULDER SYS REV INSERT 39MM 9TH B 12.5DG R; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. TORNIER FLEX SHOULDER SYS REV INSERT 39MM 9TH B 12.5DG R; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number DWF395B
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Extraskeletal ossification (1428); Loss of Range of Motion (2032); Metal Related Pathology (4530); Implant Pain (4561)
Date of Event 07/17/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The patient reported pain and decreased function, leading to a revision surgery.The reason for the revision surgery was listed as heterotopic ossification, severe stiffness, and nickel metal allergy.Components including the tray, insert, and glenosphere were explanted, while the stem, baseplate, and central screw were retained.The relatedness to the study device and the surgical procedure was assessed as possible.
 
Additional Manufacturer Narrative
The reported event was not confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.H3 other text : device disposition unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER FLEX SHOULDER SYS REV INSERT 39MM 9TH B 12.5DG R
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR   38330
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18963670
Report Number3000931034-2024-00124
Device Sequence Number12802311
Product Code KWS
UDI-Device Identifier03700434020649
UDI-Public03700434020649
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K151293
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source Study
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/06/2024
Device Catalogue NumberDWF395B
Device Lot Number7787AU
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/26/2024
Initial Report FDA Received Date03/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/07/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
Patient Weight77 KG
-
-