• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIRCRAFT MEDICAL LIMITED MCGRATH MAC; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIRCRAFT MEDICAL LIMITED MCGRATH MAC; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 350-013-000
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/22/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, the videolaryngoscope's blade anti-fog was not good enough as it was foggy before use.There was no patient involvement.
 
Additional Manufacturer Narrative
D10: concomitant products: 350-013-000, mcgrath mac 4 blade 350-013-000 50 ct (lot#:23h0035jzx).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the videolaryngoscope's blade anti-fog was not good enough as it was foggy before use.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
The date received by manufacturer that was provided in the initial report was incorrect.The correct date received by manufacturer is 3/22/2024.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: g3, g4 correction: b3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCGRATH MAC
Common Device Name
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
AIRCRAFT MEDICAL LIMITED
7 cross way
dalgety bay,fi KY11 9JE
UK  KY11 9JE
Manufacturer (Section G)
AIRCRAFT MEDICAL LIMITED
7 cross way
dalgety bay,fi KY11 9JE
UK   KY11 9JE
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18963708
Report Number3010244187-2024-00006
Device Sequence Number18134149
Product Code CCW
UDI-Device Identifier10884521824324
UDI-Public10884521824324
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2026
Device Model Number350-013-000
Device Catalogue Number350-013-000
Device Lot Number23H0035JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2024
Supplement Date Received by Manufacturer05/29/2024
10/14/2024
03/22/2024
Initial Report FDA Received Date03/22/2024
Supplement Report FDA Received Date06/03/2024
10/15/2024
10/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTE ON H11.
Patient SexUnknown
-
-