SMITH & NEPHEW, INC. TWINFIX ULTRA 5.5MM W/2 UB (WHT & BLUE); FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72202895 |
| Medical Device Problem Codes |
Break (1069); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).The reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The insertion device was returned with the distal end fractured away.The anchor and suture material were not returned.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failures and/or harms were documented appropriately, and there were no indications to suggest the anticipated risks are not adequate.The root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, attempted correction of a damaged device, or an inadvertent impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a shoulder arthroscopy, when the threading was completed, the twinfix ultra anchor disassembled from the handle and internal screw.The anchor was only one turn away from reaching the laser line and remained fixed.The procedure was successfully completed without surgical delay using the same device.No further complications were reported.
|
| |
|
Search Alerts/Recalls
|
|
|