BOSTON SCIENTIFIC CORPORATION TRIA SOFT; STENT, URETERAL Back to Search Results
Model Number M0061903430
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a040601 captures the reportable event of stent buckled material, inside the patient.
 
Event or Problem Description
It was reported to boston scientific corporation that a tria soft ureteral stent was used during a ureteroscopy procedure in the ureter performed on (b)(6) 2024.During insertion, when the stent was being inserted into the ureter using the enclosed pusher, the stent was found buckled.The procedure was successfully completed with another tria soft ureteral stent.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a040601 captures the reportable event of stent buckled material, inside the patient.Block h10: the returned tria ureteral stent was analyzed, and a visual and microscopic evaluation noted that the bladder coil buckled.Additionally, the suture and positioner were not returned.No other problems with the device were noted.The reported event was confirmed.Taking all available information into consideration, most likely, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.It is possible to conclude that this problem could be caused by operational factors.Interaction of the device between the positioner and the guide wire while the device was pushing up, such as excess of force applied over the device during the procedure could have contributed to the problem, consistently leading to device being buckled and torn.Consequently, affect the performance of the device.Therefore, all compiled information on this investigation determines that the most probable cause is "adverse event related to procedure" since the adverse event occurred during the procedure and the device had no influence on event.
 
Event or Problem Description
It was reported to boston scientific corporation that a tria soft ureteral stent was used during a ureteroscopy procedure in the ureter performed on (b)(6) 2024.During insertion, when the stent was being inserted into the ureter using the enclosed pusher, the stent was found buckled.The procedure was successfully completed with another tria soft ureteral stent.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIA SOFT
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18965202
Report Number2124215-2024-15871
Device Sequence Number7965786
Product Code FAD
UDI-Device Identifier08714729959984
UDI-Public08714729959984
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K191609
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/26/2025
Device Model NumberM0061903430
Device Catalogue Number1983-02
Device Lot Number0030429040
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/27/2024
Supplement Date Received by Manufacturer04/11/2024
Initial Report FDA Received Date03/23/2024
Supplement Report FDA Received Date05/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-