Catalog Number DWJ502
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
 
Event or Problem Description
An issue arose with a baseplate where the central screw wouldn't fully engage.Fortunately, this didn't affect the patient or surgery due to having a backup available.The incident occurred during a primary reverse surgery.There was a brief delay to resolve the issue and obtain another baseplate.
 
Event or Problem Description
An issue arose with a baseplate where the central screw wouldn't fully engage.Fortunately, this didn't affect the patient or surgery due to having a backup available.The incident occurred during a primary reverse surgery.There was a brief delay to resolve the issue and obtain another baseplate.
 
Additional Manufacturer Narrative
The reported event was confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the visual inspection was performed on returned device and revealed no obvious signs of wear or damage and spacer location is incorrect.The functional test was performed and it was observed that the reported device has two spacers.There supposed to be only one spacer assembled in the product, due to which the central screw is not able to seat on the baseplate properly.Hence, the reported event is confirmed.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on investigation, the root cause was attributed to a manufacturing related issue.Hence, the nonconformance is initiated to further access the issue.If any further information is provided, the complaint report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER PERFORM REVERSED LATERALIZED BASEPLATE 3MM DIA 25MM
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18965631
Report Number0001649390-2024-00123
Device Sequence Number15494198
Product Code PHX
UDI-Device Identifier00846832092123
UDI-Public00846832092123
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K161742
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberDWJ502
Device Lot NumberCZ3923353
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/26/2024
Supplement Date Received by Manufacturer04/24/2024
Initial Report FDA Received Date03/23/2024
Supplement Report FDA Received Date05/20/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-